| AC Voltage |
Ranges 600.0 mV, 6.000 V, 60.00 V, 600.0 V, 1000 V; Resolution 0.1 mV to 1 V; Accuracy 45 Hz to 65 Hz: ±(1.0% + 3); 65 Hz to 200 Hz: ±(4.0% + 3); 200 Hz to 500 Hz: ±(15% + 3); crest factor ≤ 3 at full scale up to 500 V; full time low pass filter; do not exceed 107 V-Hz |
| AC mV |
Range 600.0 mV; Resolution 0.1 mV; Accuracy 45 Hz to 500 Hz: ±(1.0% + 3) |
| DC Voltage |
Ranges 6.000 V, 60.00 V, 600.0 V, 1000 V; Resolution 0.001 V to 1 V; Accuracy 6 V to 600 V: ±(0.09% + 2); 1000 V: ±(0.15% + 2) |
| DC mV |
Range 600.0 mV; Resolution 0.1 mV; Accuracy ±(0.09% + 2) |
| Resistance |
Ranges 600.0 Ω, 6.000 kΩ, 60.00 kΩ, 600.0 kΩ, 6.000 MΩ, 50.00 MΩ; Resolution 0.1 Ω to 0.01 MΩ; Accuracy 600 Ω: ±(0.5% + 2); 6 kΩ to 600 kΩ: ±(0.5% + 1); 50 MΩ: ±(1.5% + 3) |
| Continuity |
Range 600 Ω; Resolution 1 Ω; Beeper on at < 25 Ω; detects opens or shorts of 600 µs or longer |
| Diode Test |
Range 2.000 V; Resolution 0.001 V; Accuracy ±(1% + 2) |
| Capacitance |
Ranges 1000 nF, 10.00 µF, 100.0 µF, 9999 µF; Resolution 1 nF to 1 µF; Accuracy 1000 nF to 100 µF: ±(1.2% + 2); 9999 µF: ±10% typical |
| AC Current (with iFlex) |
Ranges 999.9 A, 2500 A; Resolution 0.1 A and 1 A; Accuracy 45 Hz to 500 Hz: ±(3.0% + 5) |
| Frequency |
Ranges 99.99 Hz and 999.9 Hz; Resolution 0.01 Hz and 0.1 Hz; Accuracy ±(0.1% + 1) |
| Voltage Input Characteristics |
AC, DC, mV inputs: nominal input impedance > 10 MΩ, < 100 pF; common mode rejection ratio > 60 dB DC to 60 Hz (AC volts), > 120 dB at DC, 50 Hz or 60 Hz (DC volts and mV); normal mode rejection > 60 dB at 50 Hz or 60 Hz; overload protection 1100 V rms |
| Resistance / Capacitance Input |
Open circuit test voltage < 2.7 V DC; full scale voltage to 6 MΩ < 0.7 V DC; full scale voltage 50 MΩ < 0.9 V DC; typical short circuit current < 350 mA; overload protection 1100 V rms |
| Continuity / Diode Input |
Open circuit test voltage < 2.7 V DC; full scale voltage 2.000 V DC; typical short circuit current < 1.1 mA |
| MIN/MAX Recording |
AC functions: captures changes > 900 ms, 40 counts; DC functions: captures changes > 350 ms, 12 counts |
| Infrared Camera Temperature |
Range -10 °C to 200 °C; resolution 0.1 °C; centerpoint temperature measurement; accuracy ±5 °C or ±5%, whichever is greater, at 25 °C ambient; emissivity 0.95 fixed |
| Infrared Camera Optics |
Resolution 102 × 77; image capture frequency 8 Hz; detector type uncooled vanadium oxide; infrared spectral band 7.5 µm to 14 µm |
| Image Capture / Storage |
Image review before save; internal memory stores up to 100 images; image transfer via Fluke Connect or SmartView software; file format .is2 |
| Display Update Rate |
4 readings per second |
| Temperature Coefficient |
0.1 × specified accuracy per °C for temperatures below 18 °C or above 28 °C |
| Environmental Conditions |
Operating temperature -10 °C to 50 °C; storage temperature -20 °C to 60 °C; relative humidity 0% to 90% (0 °C to 35 °C), 0% to 75% (35 °C to 40 °C), 0% to 45% (40 °C to 50 °C); operating altitude 2000 m; storage altitude 12000 m |
| Safety / Compliance |
Maximum voltage between any terminal and earth ground 1000 V; measurement category CAT III 1000 V, CAT IV 600 V; certifications CSA, FCC, CE |